Goatanti-Mouse(Murine)IgGAntibody(Biotin)
詳情介紹:
Fragment | H&L |
Characteristics | This reagent is intended to be used in a labeled streptavidin-biotin immunoenzymatic antigen detection system. This technique involves the sequential incubation of the specimen with an unconjugated mouse primary antibody specific to the target antigen, a biotinylated secondary antibody which reacts with the primary antibody, enzyme- labeled streptavidin, and substrate-chromogen. |
Protocol |
1. Deparaffinize and rehydrate formalin-fixed paraffin-embedded tissue section. 2. Perform appropriate pretreatment and blocking if required. Wash 3 times in buffer. 3. Apply primary antibody and incubate according to manufacturer’s protocol. 4. Wash 4 times in buffer. Apply Biotinylated Goat Anti-Mouse and incubate 10-15 minutes at room temperature. Wash 4 times in buffer. 5. Apply appropriate enzyme-labeled streptavidin and incubate according to the manufacturer’s instructions. 6. Rinse 4 times in buffer. Apply chromogen and substrate according to manufacturer’s instructions. 7. Apply counterstain according to manufacturer’s instructions. 8. Dehydrate and coverslip or use appropriate mounting medium. |
Restrictions | For Research Use only |
Format | Liquid |
Buffer | This product contains biotinylated goat anti-mouse IgG (H+L) in PBS, pH 7.6 with less than 0.1% sodium azide. |
Preservative | Sodium azide |
Precaution of Use |
1. Take reasonable precautions when handling reagents. Wear appropriate Personal Protective Equipment. 2. Avoid contact of reagents with eyes, mucous membranes and skin. If reagents come into contact with sensitive areas, wash with copious amounts of water. 3. Consult local or state authorities with regard to recommended method of disposal. 4. Avoid microbial contamination of reagents. |
Handling Advice | Do not freeze. |
Storage | 4 °C |
Storage Comment | The user must validate any other storage conditions. The reagents must be returned to the storage conditions identified above immediately after use. When properly stored, the reagent is stable to the date indicated on the label. Do not use the reagent beyond the expiration date. There are no definitive signs to indicate instability of this product, therefore, positive and negative controls should be tested simultaneously with unknown specimens. |